NEBB
Testing, Adjusting & Balancing
Program
NEBB’s Testing,
Adjusting and Balancing Program for
air and hydronic systems provides
for certification of firms and
qualification of individuals that
meet criteria established by NEBB.
Purpose of
Certification Program
The
purpose of NEBB certification is to
offer tangible proof of competent
firms and supervisors qualified in
the proper methods and procedures
for the various categories of NEBB
certification.
Tab
Program Objectives
The
following objectives were created
for this purpose:
Suspension
A
firm's certification shall
automatically be suspended when the
firm no longer meets all of the NEBB
certification requirements. The
certification of a firm suspended
may be reinstated by NEBB at such
time as it again meets the
requirements for certification. The
period of suspension may be no
longer than one year at which time
the firm will lose its certification
automatically.
A
certified firm must notify the local
NEBB Chapter in writing within ten
days of the temporary or permanent
loss of a designated qualified
supervisor.
Loss
of Certification
Certification
may be terminated for failure of the
firm to abide by the objectives and
performance standards of NEBB.
Termination of certification by NEBB
requires vote of the Board of
Directors of NEBB provided that by
similar vote the Board shall first
find that certification is
prejudicial to the best interests of
the Bureau and provided further that
the certified firm in question shall
have had opportunity upon written
notice of at least fifteen (15) days
to show cause why the certification
should not be terminated.
INSTRUMENTS
FOR TAB CERTIFICATION
OWNERSHIP
REQUIREMENTS
It
is the responsibility of the NEBB-certified TAB Firm to provide
appropriate instrumentation that meets
the minimum requirements as called for
in the 7th Edition NEBB
Procedural
Standards for Testing Adjusting
Balancing of
Environmental Systems
TABLE 4-1 for use on a
project. Instrumentation used on a
NEBB project shall be in proper
operating condition and shall be
applied in accordance with the
manufacturer’s recommendations. TABLE
4-1 lists the minimum instrumentation
that a NEBB-certified TAB Firm shall
own and maintain.
A firm certified in
one discipline is only required to own
instruments appropriate for their
category of certification as noted
above. Specifically, “A” =
instrumentation required for Air
Certification. “H” = equals
instrumentation required for Hydronic
certification.
Click
here
for Table 4-1 - Table of Required
Instrumentation. Requires
Adobe Reader. Download
it free below.
RANGE
AND ACCURACY
A
NEBB-certified TAB Firm shall possess
instruments -- of the firm's choice --
for each function and range listed in
Table 4-1. Each instrument shall have
been specifically designed to meet the
criteria (Minimum Accuracy, Range, and
Resolution) of the function.
Instrumentation
with multiple capabilities shall be
accepted for more than
one function when submitting
documentation for a firm’s
certification, providing that each
separate function meets NEBB
requirements.
Information and data
regarding accuracy of all submitted
instrumentation for the stated
functions shall be available from the
manufacturer.
The accuracy and range
as reported by the instrument
manufacturer shall be verified by a
testing laboratory traceable to the
National Institute of Standards and
Technology or equivalent institute in
countries other than the United
States. Calibration requirements for
each function are specified and shall
be met. Some instruments such as
U-tube manometers and inclined
manometers may not require
calibration. However, if a
"mechanical/electrical" device is
substituted or employed in place of
these types of instruments, the
indicated calibration requirements
noted shall apply.
Firms with multiple
sets of instrumentation shall comply
with either of the following
conditions as a minimum requirement
for NEBB certification:
a) Calibrate all
instrumentation used by the firm on
TAB projects in accordance with
Table 4-1.
or
b) Maintain a
complete set of calibrated
instrumentation used for comparison
with regularly used instrumentation.
Periodic checking of regularly used
instrumentation against the calibrated
set shall be performed.
Acceptance criteria for the results of
the comparisons are the
responsibility of the NEBB Qualified
TAB Supervisor.
Supervisors
must understand the importance of
using accurate instrumentation in the
field, and shall be prepared to have
witnesses verify their work with the
Firm’s calibrated set of
instrumentation. Results of the data
verification shall validate the
accuracy of the instrumentation used
to perform the work.
APPROPRIATE
APPLICATION OF INSTRUMENTS
Instruments shall be
used in accordance with manufacturer’s
recommendations. The most suitable
instrument, or combination of
instruments, should be employed for a
particular measurement or reading.
For example, a traverse may be
accomplished with a Pitot tube and
manometer (digital, analog, or
incline); it is not acceptable to use
a Pitot tube with another device that
does not provide the same overall
accuracy.